Immunohistochemistry: An Overview + Steps to Better IHC Staining

Email Subscription

Be the first to know about upcoming webinars and join the hundreds that enhance their knowledge with our new videos, articles, and other industry leading content.

Subscribe

Geoffrey O. Rolls is a Histology Consultant with decades of experience in the field. He is a former Senior Lecturer in histopathology in the Department of Laboratory Medicine, RMIT University in Melbourne, Australia.

Steven Westra is a renowned antibody staining consultant with over 10 years of experience in the immunohistochemistry industry.

James Anderson, Global Marketing Manager Leica Biosystems has over 17 years of experience in the Anatomic Pathology industry.

An Introduction to Immunohistochemistry

Immunohistochemistry (IHC) is used in histology to detect the presence of specific protein marker that can assist with accurate tumor classification and diagnosis. This guide illustrates the basic steps used to create an IHC stain.

An Introduction to IHC Staining

Immunohistochemistry (IHC) is used in histology to detect the presence of specific protein markers that can assist with accurate tumor classification and diagnosis. Immunohistochemistry has evolved to complement the Hematoxylin & Eosin (H&E) and Special Stain techniques that typically show tissue morphology (structure). Where H&E and Special Stains are non-specific, IHC is directed to a specific protein marker or markers. IHC is used as a diagnostic tool to assist in the diagnosis of solid tumors and cytological specimens and has been used as a mainstream diagnostic tool for almost half a century.

Target Antigen

Antigens are proteins that are within or on the surface of a cell. Pathologists look for the presence or absence of particular antigens to assist with diagnosis.

There are many hundreds of antigens that have been found to be diagnostically useful. Often, a pathologist will use a “panel” of multiple antigens to help fully classify a particular tumor.

Primary Antibody

The first stage of IHC is the application of a primary antibody that binds specifically to the target antigen.

There are two main types of antibody, polyclonal and monoclonal. Polyclonal antibodies have an affinity with, and bind to, multiple epitopes (or parts) or the target antigen, and as such are more prone to cross-react to non-target antigens.

Monoclonal antibodies have an affinity to only one epitope and tend to produce, cleaner, more specific staining but are less sensitive or intense.

Secondary Antibody

Next, secondary antibodies bind to the primary antibody. This is known as indirect IHC. It is now commonly used as multiple secondaries can bind to a single primary to amplify the staining intensity.

The Detection System

The detection system builds on the secondary. Modern chromogenic detection utilizes enzymes such as Horseradish Peroxidase (HRP) that are conjugated (joined) to an antibody. Multiple enzymes attached to the antibody are known as polymers, and they again produce more intense staining as there are more molecules for the chromogen to attach to.

The Chromogen

Finally, a substrate forms an insoluble colored precipitate that can be visualized under a microscope. There are two commonly used chromogens: DAB (brown) or AP (red)

DAB is used for most applications as it provides strong and permanent stains. AP Red (or another red chromogen) is used mainly for skin sections where the brown DAB may be masked by brown melanin pigment.

Both DAB and AP red are sometime used on the same tissue section to allow the pathologist to visualize two antigens on the one side. This is a process known as double staining.

A Completed IHC Stain

This is a typical IHC stain where the brown precipitate indicates the presence of the target antigen – in this case, Cytokeratin 5 on a prostate biopsy. The amount of staining, the staining pattern and the location of staining (cell cytoplasm nucleus or membrane) all provide information for the diagnosing pathologist.

The blue background is a hematoxylin counter-stain that is often applied after the chromogen. The counter-stain provides a contrast to the chromogen and also helps the pathologist visualize the underlying tissue structure.

15 Steps to Better IHC

From patient to pathologist, preparing tissue specimens for histological examination requires care, skill and sound procedures. This guide provides practical advice on best-practice techniques and simple ways to avoid common errors.

Tips for better IHC are highlighted in this section. We hope each step provides a valuable reminder of good histology practice and helps with troubleshooting when unacceptable results do occur.

Want to see all 101 Steps to Better Histology?

Click here to download your copy today!

 

Step 1 - Use High Quality Sections

 Take particular care to use thin, flat sections that have been thoroughly dried onto the slide. Preferably use charged slides or APES coated slides for IHC.

 Uneven, poorly-adhering sections stain unevenly with variable background staining.

 

 

Section A: A bubble under the section (from mounting) has resulted in subsequent detachment of the section during staining (tonsil, CD45).
Section B: B A poor quality section, that has not been properly flattened and dried before staining, has lifted making the slide unsatisfactory (tonsil, CD3).

Step 2 - Ensure Optimal Fixation

 Good quality fixation using known and consistent fixation conditions (fixative type, pH, temperature, time) produces the best results. Specimens should be checked prior to processing to determine if further fixation is required.

 Inconsistent fixation conditions, producing under-fixed or over-fixed tissues, produce variable results and make troubleshooting difficult.

Uneven fixation (zonal fixation) has resulted in uneven staining in this section (breast tumor, ER).

Step 3 - Avoid Section Adhesion Problems

 Avoid the use of protein-based section adhesives in the flotation bath (glue, starch, or gelatin), particularly on charged slides.

 Protein-based adhesives can block the surface of the charged slide. This causes inconsistent adhesion and leads to uneven staining due to pooling of IHC reagents beneath lifting sections.

A line of thick protein-based section adhesive has stained adjacent to the section (breast, PR).

Step 4 - Avoid Concentration Gradients

 Concentration gradients are avoided by careful application of reagents.

 “We sometimes see strong staining at one end of the slide progressing to weak staining at the other.”

Sections A and B are micrographs taken from opposite ends of the same slide. One end of the slide shows strong staining (A) whereas at the other end (B) the staining was very weak. This is an extreme example of a concentration gradient created during staining (tonsil, CD45).

Step 5 - Choose Antibody Carefully

 Choose your primary antibody carefully with regard to its sensitivity and specificity. Be aware that antibodies sold by different suppliers often come from the same source and are repackaged/branded for sale. It is important to use the clone name when assessing an antibody.

 “We buy our antibodies based on price alone.”

These sections of human tonsil from the same block have been stained with the B cell marker CD20 using primary monoclonal antibodies from different sources (suppliers). In each case the recommended pretreatment and optimized dilution was used. There is an obvious difference in the quality of the results achieved.

Step 6 - Read Specification Sheets

 Know your primary antibody. Always check the specification sheet to determine the suitability of your method for a particular antibody. Specification sheets should be updated when a new batch of antibody is purchased.

 “We don’t have access to the antibody specification sheets in our laboratory.”

These sections of intestine have been stained for Cytokeratin AE1/AE3. Different retrieval conditions were used for each section. Section A shows unacceptable weak staining, while section B shows strong precise staining.

Step 7 - Optimize Retrieval Methods

 Choose appropriate unmasking conditions for the primary antibody being used, the tissue being stained, and the fixation employed (pH, reagent, reaction conditions).

 The same retrieval technique is used for all primaries on the assumption that there is a successful universal HIER method.

Prostate sections stained for Cytokeratin 34βE12. Section A shows weak staining while section B is stronger and sharper. The only difference between the two was the retrieval method used.

Step 8 - Consider Antibody Cross-reactivity

 Be aware of any potential problems with antibody cross- reactivity (read the specification sheet).

 No attempt is made to explain unexpected positive staining.

Palatine tonsil showing the base of a tonsillar crypt stained for CD5, a lymphocyte marker that stains mainly T cells. This particular clone (4C7) cross- reacts with epithelial cells deep in the crypt.

Step 9 - Block Endogenous Peroxidase

 For peroxidase-based detection systems, always use a peroxidase-blocking step.

 Non-specific staining is often seen in erythrocytes, granulocytes, monocytes, and in muscle. This is due to incompletely-blocked endogenous peroxidase.

Spleen showing typical, non-specific staining of erythrocytes due to incomplete blocking of endogenous peroxidase. Here the natural peroxidase present in the red cells has reacted with the DAB chromogen.

Step 10 - Avoid Background Staining

 Appropriate protein block is always used.

 Generalized background staining is sometimes seen due to ineffective protein block.

Normal tonsil stained for Kappa light chain showing a heavy back-ground stain due to ineffective protein block.

Step 11 - Use an Appropriate Detection System

 Choose an appropriate detection system that will provide precise, specific staining with adequate sensitivity.

 “We have been using the same detection system for a long time and see no reason to change. Sometimes our stains are weak and are not as sharp as we would expect.”

Sections A and B are from the same specimen but have been stained using different detection systems. Note the difference in the intensity and precision of the stains. (tonsil, CD20).

Step 12 - Standardize Washing Steps

 Use standardized washing steps throughout (duration, volume and form of agitation). This will ensure consistency of results.

 Results are very variable within runs with the same antibody and between runs on different days. This can be due to different washing techniques used by different operators.

Sections A and B are from the same specimen and have been stained manually using the same reagents. Note the difference in the level of background staining in the stratified epithelium. This is probably due to a difference in the efficiency of the washing technique used. (tonsil, CD20).

Step 13 - Optimize Counterstaining

 The level of nuclear counterstain is carefully regulated and standardized so as not to obscure positive staining. The counterstain should provide the best possible contrast between chromogen and background tissue elements. An appropriate counterstain is chosen for the chromogen used.

 Nuclear counterstain is sometimes very strong. This can obscure weak specific staining.

Tonsil stained for Ki67, a nuclear marker for proliferating cells. The two sections are from the same specimen showing different levels of hematoxylin counterstaining. Slide A shows staining that is too strong and would obscure a weak positive reaction. Slide B shows a better level of staining.

Step 14 - Use Appropriate Controls

 Always use appropriate positive and negative controls that are carefully examined to validate results. Internal positive and negative controls are also important and provide an excellent means of ensuring quality assurance in IHC.

 “We only do controls when our method doesn’t seem to work. If we did them for every run people wouldn’t bother to look at them.”

Tonsil stained for Ki67. This was the negative control slide and the nuclei should not be stained. The primary antibody was mistakenly applied to this slide instead of the negative control reagent.

Step 15 - Evaluate Results Carefully

 Know what to look for and where to look when evaluating your test sections and controls after staining.

 If staining is observed in test sections, it is assumed the stains are satisfactory.

Intestine stained for AE1/AE3. Unexpected weak staining of the crypt epithelium has occurred. On investigation it was found that CK20 had been wrongly used as the primary antibody.

Bonus tip: Choosing the Right Antibodies

A simple, but sometimes overlooked step is to choose antibodies that work for immunohistochemistry. This can save you a lot of headaches down the road. Search for specific antibodies from literature, vendors, as well as your peers.

There are several things to keep in mind with antibodies. What might work well in one laboratory might not be optimal for your laboratory. Each antibody needs to be tested with your staining system. Antibodies overtime can lose their staining intensity. Exposure to air and light can cause this.

Concentrates vs. RTU antibodies

When selecting antibodies, there are two main options to consider; a concentrated format or a pre-diluted, Ready-To-Use (RTU) format. Concentrates are flexible, have a lower initial purchase price, and can generally be used within any staining system, both automated and manual, subject to the manufacturer’s recommendations. The working dilution of concentrates can be optimized to balance cost, staining time and quality. Given the large working dilution range, concentrates can be varied at any time to accommodate changes to laboratory practice or for multiple protocols for a particular antibody. However, concentrates do require preparation time and validation. As there is no definitive way of determining the properties and stability of a diluted antibody without well controlled and executed studies, staining quality may be compromised as subtle deteriorations may not be noticed. The advantages of RTUs include increased laboratory efficiency, better quality control and easier reagent management. They eliminate time spent on working dilution, preparation and the time to validate the assay. Consistency is enhanced with run-to-run variation reduced, especially in conjunction with automated stainers and associated detection systems. With a defined number of tests and manufacturer-verified expiry, RTUs simplify antibody management. Additionally, RTUs can contribute to laboratory growth by making it easy to adopt new antibody assays as the amount of validation work is greatly reduced.

Bonus tip: Pre-staining standardization is critical

Any pathologist, lab manager or histotechnologist will readily acknowledge that preparation for IHC staining begins the moment tissue is acquired. The literature documents optimal conditions for tissue fixation, processing and sectioning to ensure morphology and antigenicity are maintained. Further improvements in maintaining consistency to correctly control for these factors could be pursued. This may involve a laboratory establishing itself at the site of collection, recognizing that accessioning and sample preparation begins here.

Automated accession, LIS and sample tracking infrastructure should extend out of the laboratory and into surgery/point of biopsy, enabling standardized prep-equipment (”controlled” pots, fixatives) to be monitored for optimal processing downstream. A parallel industry that recognized this value is the blood analysis laboratory, which utilizes standardized coated collection vials with bar coded patient information for tracking and accessioning.

Linking fixation, tissue processing and IHC staining will add value to control quality. Laboratories that can monitor and record fixation and tissue processing conditions and link these with IHC staining protocols will be able to report a diagnosis under a tighter quality controlled environment. The frequent use of controls for pre processing variability (eg. control every 100 slides) may also provide a good measure for staining instrument tolerances and performance. These workflows and technologies will evolve as the IHC industry matures. However there are a number of good practices that are known to establish and maintain high quality and consistent results from IHC staining (Table).

Process Location Common Issues that Impact IHC
Excision/Biopsy Surgery/Physician Office »» Mechanical tissue damaged by excision equipment
»» Margins not clean or area of interest missed
»» Ischemia and cell death due to delay between excision and fixation, as well as surface area to fixative ratios
»» Mislabeling of samples
»» Over-fixation
Accessioning Laboratory »» Sample mislabeling
»» Overlooked samples
Grossing Laboratory »» Slices too thick
»» Specimen trauma caused by incorrect handling
»» Cross-contamination with other cases
»» Inappropriate cassette type
»» Overloaded cassettes
Tissue processing Laboratory »» Poor decalcification (where required)
»» Incompete fixation or over-fixation (see Figure 2)
»» Inappropriate schedule chosen
»» Poor quality reagents
»» Dirty reagents
Embedding Laboratory »» Incorrect orientation
»» Inappropriate or overfilled mold
»» Mechanical trauma due to rough handling or excessive heat
»» Inappropriate cassette type
»» Overloaded cassettes
Microtomy Laboratory »» Poor quality knife or incorrect knife profile
»» Incorrect knife tilt angle
»» Damage due to freezing
»» Rushed sectioning causing "chatter"
Flotation Laboratory »» Dirty or contaminated water
»» Cross-contamination caused by incorrect bath cleaning or floating-out sections for multiple blocks
»» Incorrect water temperature
»» Over-expanded sections
»» Bubbles forming under the tissue or sections lifting
»» Time before staining commences

Click here to download the full 101 Steps to Better Histology

This reference document is presented as a service to health care professionals by Leica Biosystems and has been compiled from available literature. Although every effort has been made to report faithfully the information, Leica Biosystems cannot be held responsible for the correctness. This document is not intended to be, and should not be construed as medical advice. For any use, the product information guides, inserts and operation manuals of the various drugs and devices should be consulted. Leica Biosystems and the editors disclaim any liability arising directly or indirectly from the use of drugs, devices, techniques or procedures described in this reference document.