Mirth Connect security advisory
2026 November
CVE-2026-82583, CVE-2026-78224, CVE-2026-82578
Last updated: 2026 November 13th
Background
Leica Biosystems (LBS) has been notified of three security vulnerabilities affecting Mirth Connect version 4.7.1 and earlier: CVE-2026-82583, CVE-2026-78224, and CVE-2026-82578. Mirth Connect is used in certain Leica Biosystems products to facilitate communication with Laboratory Information Systems (LIS).
Leica Biosystems has assessed its products and identified multiple products that contain affected versions of Mirth Connect. The impact and recommended actions vary by product and configuration, as described in the table below. Leica Biosystems is not aware of any exploitation of these vulnerabilities affecting Leica Biosystems products.
Leica Biosystems recommends that customers of affected products review the product-specific guidance below, assess their network configuration, and implement the recommended mitigations where applicable.
Leica Biosystems has determined that its products do not use Mirth Connect in configurations that permit exploitation of CVE-2026-78224 or CVE-2026-82578. The product-specific guidance below therefore focuses on CVE-2026-82583 and provides information on affected products and the actions customers should take to mitigate the associated risk.
Leica Biosystems is currently unable to provide an update to the Mirth Connect component. Customers should therefore follow the product-specific recommendations below to mitigate CVE-2026-82583.
Product Status
| Product | Status Regarding Mirth Connect Vulnerability | Recommendations and Comments |
|---|---|---|
| Aperio AT2, Aperio AT2 DX | Not Vulnerable | Mirth Connect software is not present. |
| Aperio Bridge Data Broker | Not Vulnerable | Mirth Connect software is not present. |
| Aperio CS2 | Not Vulnerable | Mirth Connect software is not present. |
| Aperio CS5, Aperio CS5 DX | Not Vulnerable | Mirth Connect software is not present. |
| Aperio eSlide Manager | Not Vulnerable | Mirth Connect software is not present. |
| Aperio eSlide Manager with LIS Connectivity | Potentially Vulnerable |
Mirth Connect is a component of the LIS Connect feature used to integrate eSlide Manager with a Laboratory Information System (LIS). The security guidance provided with eSlide Manager recommends that the Windows operating system hosting the eSlide Manager software be configured to deny network access to all ports other than those explicitly documented as required for operation. Correct implementation of this guidance blocks network access to the Mirth Connect administrative API on TCP port 8443, thereby preventing exploitation of CVE-2026-82583 from the laboratory network. Leica Biosystems recommends that customers verify that TCP port 8443 on the eSlide Manager Windows host is not accessible from the laboratory network. If TCP port 8443 is accessible from the laboratory network, corrective action should be taken to block network access to this port. |
| Aperio GT Platform | Not Vulnerable | Mirth Connect software is not present. |
| Aperio GT 180 RUO, Aperio GT 180 DX, Aperio GT 450 RUO, Aperio GT 450 DX, Aperio Elite RUO | Not Vulnerable | Mirth Connect software is not present. |
| Aperio ImageScope, Aperio ImageScope DX | Not Vulnerable | Mirth Connect software is not present. |
| Aperio iQC RUO, Aperio iQC DX | Not Vulnerable | Mirth Connect software is not present. |
| Aperio LV1 | Not Vulnerable | Mirth Connect software is not present. |
| Aperio Scanner Administration Manager (SAM) RUO, Aperio Scanner Administration Manager (SAM) DX | Not Vulnerable |
Affected versions of Mirth Connect may be present in Aperio Scanner Administration Manager (SAM) software versions 1.0.0+, 1.1.0+, 1.2.0–1.2.2, and 1.3.0–1.3.2. However, Aperio SAM does not use or expose the Mirth Connect functionality associated with these three vulnerabilities. The vulnerable functionality is not reachable through the product’s Mirth configuration; therefore, Aperio SAM is not affected. No action is required to address these vulnerabilities; however, Leica Biosystems recommends upgrading to the latest available version of Aperio SAM to obtain the most current security updates. |
| Aperio VERSA | Not Vulnerable | Mirth Connect software is not present. |
| Aperio WebViewer, Aperio WebViewer DX | Not Vulnerable | Mirth Connect software is not present. |
| BOND-ADVANCE, BOND Controller | Not Vulnerable | Mirth Connect software is not present. |
| BOND-III | Not Vulnerable | Mirth Connect software is not present. |
| BOND-MAX | Not Vulnerable | Mirth Connect software is not present. |
| BOND RX, BOND RXm | Not Vulnerable | Mirth Connect software is not present. |
| CEREBRO | Potentially Vulnerable |
CEREBRO software version 3.2 and earlier includes Mirth Connect installed on the CEREBRO Server. The CEREBRO Server configures the Windows Firewall to block network access to the Mirth Connect administrative API on TCP port 8443, thereby preventing exploitation of CVE-2026-82583 from the network. For additional assurance, Leica Biosystems recommends that customers verify that TCP port 8443 on the CEREBRO Server is not accessible from the network. If TCP port 8443 is accessible, customers should take corrective action to restore the firewall configuration and block network access to this port. |
| CytoVision, CytoVision DX, CytoInsight GSL | Not Vulnerable | Mirth Connect software is not present. |
| HistoCore Arcadia C | Not Vulnerable | Mirth Connect software is not present. |
| HistoCore Arcadia H | Not Vulnerable | Mirth Connect software is not present. |
| HistoCore PEARL | Not Vulnerable | Mirth Connect software is not present. |
| HistoCore PEGASUS (PLUS) | Not Vulnerable | Mirth Connect software is not present. |
| HistoCore PELORIS 3 | Not Vulnerable | Mirth Connect software is not present. |
| HistoCore SPECTRA CV | Not Vulnerable | Mirth Connect software is not present. |
| HistoCore SPECTRA ST | Not Vulnerable | Mirth Connect software is not present. |
| HistoCore SPRING ST | Not Vulnerable | Mirth Connect software is not present. |
| Leica ASP200 (S), Leica ASP300 (S) | Not Vulnerable | Mirth Connect software is not present. |
| Leica ASP6025 (S) | Not Vulnerable | Mirth Connect software is not present. |
| Leica CV5030 | Not Vulnerable | Mirth Connect software is not present. |
| Leica IP C | Not Vulnerable | Mirth Connect software is not present. |
| Leica IP S | Not Vulnerable | Mirth Connect software is not present. |
| Leica ST4020 | Not Vulnerable | Mirth Connect software is not present. |
| Leica ST5010 | Not Vulnerable | Mirth Connect software is not present. |
| Leica ST5020 | Not Vulnerable | Mirth Connect software is not present. |
| Leica TP1020 | Not Vulnerable | Mirth Connect software is not present. |
| PELORIS, PELORIS II | Not Vulnerable | Mirth Connect software is not present. |
| PathDX | Not Vulnerable | Mirth Connect software is not present. |
| ThermoBrite Elite | Not Vulnerable | Mirth Connect software is not present. |