Skip to main content
Knowledge-Pathway-Hero-image

Gaining Efficiencies with an Anti-Static Waste Tray on Histocore Rotary Microtomes

The cleaning process of a paraffin tray during microtomy may seem like a mundane and ordinary task. However, when you consider the time it takes to clean the tray and the number of times per day this occurs, it can add up to a significant amount of time and resources each day. Leica Biosystems’ new anti-static microtome paraffin tray was designed to resist the adhesion of paraffin sections to the tray and allow for ease of cleaning. The following research was conducted to measure the different time measurements between the following methods: cleaning a standard paraffin tray; cleaning a standard paraffin tray with wet paper towels; and cleaning the anti-static tray.

Image
paraffins-tray
Fig 1: Paraffin's tray

Each method started by filling the paraffin tray of the HistoCore AUTOCUT Automated Rotary Microtome with 7000, 4-micron sections. The tray was then emptied by first knocking it on the side of the waste bin 3 times to remove all loose paraffin. The timer was started when the remaining paraffin that was adhered to the tray was swept away with a brush and stopped when this process was completed. Each method: standard tray, standard tray with paper towel and antistatic tray, was duplicated 15 times. The mean data and statistical data is shown at right.

Results (mean):

  • Standard tray: 16.07 seconds
  • Standard with wet paper towels: 11.33 seconds
  • Anti-Tray 4.8 seconds
Image
70-reduction
Fig 2: 70% Reduction in total cleaning time from the standard tray

From the observed results, a 70% reduction in total cleaning time from the standard tray was calculated. The time savings can be further demonstrated by the number of times a tray is cleaned each day. If, for example, 10 standard trays are cleaned 10 times in an 8-hour shift, the resulting time would be 1607 seconds, or 26.8 minutes per 8 hours. If the anti-static tray is used, the resulting cleaning time would be 8 minutes per 8 hours. This is a significant difference in time, freeing up 18.8 minutes every 8 hours, or 56.4 minutes during a 24 hour day. This results in a savings of nearly 1 hour of tray cleaning time every day.

Image
Savings
Fig 3: Savings of 18.8 minutes per 8 hrs & Savings of 56.4 minutes per 24 hrs

Projections and Realized Results are specific to the institution where they were obtained and may not reflect the results achievable at other institutions.


Leica Biosystems 콘텐츠는 Leica Biosystems 웹사이트 이용 약관의 적용을 받으며, 이용 약관은 다음에서 확인할 수 있습니다. 법적고지. 라이카 바이오시스템즈 웨비나, 교육 프레젠테이션 및 관련 자료는 특별 주제 관련 일반 정보를 제공하지만 의료, 규정 또는 법률 상담으로 제공되지 않으며 해석되어서는 안 됩니다. 관점과 의견은 발표자/저자의 개인 관점과 의견이며 라이카 바이오시스템즈, 그 직원 또는 대행사의 관점이나 의견을 나타내거나 반영하지 않습니다. 제3자 자원 또는 콘텐츠에 대한 액세스를 제공하는 콘텐츠에 포함된 모든 링크는 오직 편의를 위해 제공됩니다.

모든 제품 사용에 다양한 제품 및 장치의 제품 정보 가이드, 부속 문서 및 작동 설명서를 참조해야 합니다.

Copyright © 2024 Leica Biosystems division of Leica Microsystems, Inc. and its Leica Biosystems affiliates. All rights reserved. LEICA and the Leica Logo are registered trademarks of Leica Microsystems IR GmbH.