Skip to main content
Knowledge-Pathway-Hero-image

A single BOND-PRIME can process the work of two BenchMark ULTRA instruments.1

A laboratory considering changing their IHC/ISH workflow from BenchMark ULTRAs to BOND-PRIMEs wanted to understand the gains that BOND-PRIME would deliver to their laboratory. Knowing that a comparative run between a single BOND-PRIME and a single ULTRA showed that a BOND-PRIME would complete slides faster, the lab wanted to see if a single BOND-PRIME could deliver the throughput of two of their ULTRA instruments on a standard workday. The lab selected archival staining run reports from two of their ULTRA instruments on a single day that represented average slide loading numbers and included single DAB, single RED and double sequential staining protocols. The slide loading data was consolidated into a single run list. A replicate of these slides was run on a single BOND-PRIME instrument with slides loaded at the same time points in a like for like continuous mode. This test was then repeated with two different ULTRA instruments reports representing different protocol usage scenarios.

In a repeatable outcome, two tests demonstrated that one BOND-PRIME instrument could achieve the same slide numbers in the same amount of time as two BenchMark ULTRA instruments when run in a continuous mode as per this laboratory’s IHC workflow.

TEST A – 107 slides – 10.5 hours
TEST B – 113 slides – 10.5 hours
Two BenchMark ULTRAs
One BOND-PRIME

Image
Picture1.png

TEST A

Test A consisted of a combination of two BenchMark ULTRA runs with a total of 107 slides processed. These slides included the following protocols:

  • 79 IHC DAB chromogen slides
  • 23 IHC Red AP chromogen slides
  • 5 IHC sequential double stained slides with both DAB and Red AP chromogens

 

Image
Picture2.png
Image
Picture3.png
Image
Picture4.png

TEST B

Test B consisted of a combination of two BenchMark ULTRA runs with a total of 113 slides processed. These slides included the following protocols:

  • 105 IHC DAB chromogen slides
  • 8 IHC sequential double stained slides with both DAB and Red AP chromogens
Image
Picture5.png
Image
Picture6.png
Image
Picture7.png

Summary of Potential Gains

Evaluating the comparative results of TEST A, and TEST B multiple benefits that this laboratory could realize can be determined.

  • In this continuous slide loading workflow, a single BOND-PRIME completed the work of two BenchMark ULTRAs in the same amount of time giving this laboratory the opportunity to change their IHC/ISH lab to BOND-PRIME instruments at a ratio vs BenchMark ULTRA of 1:2.1
  • A single BOND-PRIME instrument will regularly complete more slides earlier in the day than the two BenchMark ULTRA instruments even when processing slides at a ratio of 1:2.2
  • When continuously processing 100+ slides in this laboratory’s workflow with a 1:2 ratio BOND-PRIME vs BenchMark ULTRA, 100% of the BenchMark ULTRA slides TAT is greater than two hours, while 80% of the BOND-PRIME slides TAT is under two hours.1
  • Changing the laboratory IHC/ISH workflows to a BOND-PRIME system at a ratio of 1:2 vs the previous BenchMark ULTRA system will deliver savings to this laboratory,
    • Floor space
      • BOND-PRIME dimensions = 1217(w) x 831(d) mm3
      • BenchMark ULTRA dimensions = (1118(w) x 841(d) mm) x 24
    • Waste volume and waste container management
      • BOND-PRIME average waste volume = 57mLs per DAB IHC slide5
        • Separated into
          • 19mL Standard waste.
          • 38mL Hazardous waste.
        • 4.5L waste containers require emptying every 200 slides.5
      • BenchMark ULTRAs average waste volume = 165mLs per DAB IHC slide6
        • Single waste container with no separation of components.4
        • Quarterly decontamination procedures generate up to a further 40L of waste.4
        • 20L waste carboys require emptying every 90 slides.4
    • Instrument maintenance
      • BOND-PRIME IFU recommended maintenance, not including daily reagent top-ups, is 6 hours per year.5
      • BenchMark ULTRA IFU recommended maintenance, referring just to the quarterly decontamination procedures, state there is 10 hours of instrument maintenance per year, per instrument associated to this task alone.4
    • It can be strongly hypothesized that laboratory resource requirements can also be reduced across multiple activities,
      • Hands-on touch points for one instrument vs two would see a general reduction.
      • Instrument QC and validation for one instrument vs two would require less resource time and use of consumables.
      • Instrument QC documentation for one instrument vs two would be halved.

FOR IN VITRO DIAGNOSTIC USE

1These results are applicable for the Test A & B scenarios as described here for this laboratory and cannot be used as a standalone claim. Individual results may vary.

2As referenced in Tests A & B, in Case Study 230327 Rev A_BOND-PRIME completes slides earlier with less hands-on time and in Case Study 220218 Rev C_ Comparing the throughput of the BOND-PRIME IHC system to a traditional IHC platform

391.7500.501 BOND-PRIME User Manual

4750-600 BenchMark ULTRA Advanced Staining System Operator Guide

5220856 Rev B_BOND-PRIME Maintenance and Bulk Reagent Management Flyer

6190374 Rev D_BOND-III proves superior to BenchMark ULTRA Flyer


Related Content

Leica Biosystems 콘텐츠는 Leica Biosystems 웹사이트 이용 약관의 적용을 받으며, 이용 약관은 다음에서 확인할 수 있습니다. 법적고지. 라이카 바이오시스템즈 웨비나, 교육 프레젠테이션 및 관련 자료는 특별 주제 관련 일반 정보를 제공하지만 의료, 규정 또는 법률 상담으로 제공되지 않으며 해석되어서는 안 됩니다. 관점과 의견은 발표자/저자의 개인 관점과 의견이며 라이카 바이오시스템즈, 그 직원 또는 대행사의 관점이나 의견을 나타내거나 반영하지 않습니다. 제3자 자원 또는 콘텐츠에 대한 액세스를 제공하는 콘텐츠에 포함된 모든 링크는 오직 편의를 위해 제공됩니다.

모든 제품 사용에 다양한 제품 및 장치의 제품 정보 가이드, 부속 문서 및 작동 설명서를 참조해야 합니다.

Copyright © 2026 Leica Biosystems division of Leica Microsystems, Inc. and its Leica Biosystems affiliates. All rights reserved. LEICA and the Leica Logo are registered trademarks of Leica Microsystems IR GmbH.